No, the Detachable Tips are for single use only.
The Detachable Tips are sold sterile and ready to use. They are for single use only and must be discarded after the procedure.
You may clean the Fiber Optic Cable with a pre-steam autoclave, the STERIS® or the STERRAD® system. Use the "STERRAD® 100 System Operator's Manual" when sterilizing the fiber optic cable with a STERRAD® System. Do not use synthetic detergents or oil bases soaps as they can damage the Fiber Optic Cable.
STERRAD® is a registered trademark of Advanced Sterilization Products.
No, do not soak the Fiber Optic Cable in any disinfectants as the material will absorb the chemicals.